德國菲希爾*新鐵素體含量檢測儀FMP30
The rugged FERITSCOPE® FMP30 with plug-in type smart probe is suited for fast, non-destructive and accurate measurement of ferrite content on-site or in the lab according to the magnetic inductive method.
鐵素體含量檢測儀FMP30The measurement range is from 0.1 to 80 % Fe or from 0.1 to 110 FN. The measurement options are either measurable in Ferrit-percentage %Fe or in Ferrit numbers FN.
1,鐵素體含量檢測儀FMP30應(yīng)用:
Applications
Measurement of the ferrite content in
austenitic steel welds (of tubing etc.)
normal construction steel with austenitic chrome alloy steel welded cladding (e.g., in boilers, vessels, etc.)
duplex steel.
This easy-to-use instrument has an USB interface for a PC and printer connection. It offers memory for up to 20000 measurements with up to 100 applications. Further a detailed statistical and graphical evaluation is possible.
The calibration of the FERITSCOPE® FMP30 is traceable to internationally approved IIW secondary calibration standards. The instrument therefore meets all requirements of ANSI/AWS A4.2 and DIN EN ISO 8249 as well as of the "Basler" standard.
The picture shows one of the 3 sets of available calibration standards with certificate for different Ferrite-Number ranges.
All Fischer calibration standard sets are traceable to internationally approved IIW secondary calibration standards from TWI/England.
2,鐵素體含量檢測儀FMP30參數(shù)如下:
Features
Memory for up to 20,000 readings
Up to 100 calibratable applications
Capability of allocating readings into up to 4,000 Blocks
Date and time stamp for blocks
Statistics display of common characteristic values in the block and final results. Output of characteristic variance-analytical values
Graphical measurement display as a histogram with a Gaussian plot
Capability of entering process tolerance limits and computation of the associated process capability indices cp and cpk
Audible and visual warning when tolerance limits are exceeded
Free-running display with additional presentation of the reading as an analog bar between the tolerance limits
External key-triggered measurement acquisition, e.g., in hollow cylinders with small diameters
Capability to enable matrix measurement mode for connected multi-point measurements
Capability of averaging measurement data: Only the mean value of several readings will be stored
Measurement acquisition through area measurement possible: Only the single readings until probe lift-off are captured and averaged
Capability to measure continuously with the probe placed on the specimen
Outlier rejection for the automatic elimination of erroneous measurements
Correction of any stored reading
Application linking mode: Ability for common normalization/calibration of applications
Designations for applications through the optional PC program MP-Name
USB port to a printer
涂裝 相關(guān)儀器:涂層測厚儀,漆膜硬度測試儀,電解膜厚儀,黏度計/粘度計,X射線測厚儀,附著力測試儀,微型光澤儀,分光色差儀,霧影儀,反射率測定儀,光譜儀,涂膜干燥時間記錄儀,標準光源箱