NA23雷電沖擊峰值電壓表 NA23 Lightning impact peak voltage meter
產(chǎn)品別名
NA23型沖擊峰值電壓表,沖擊脈沖峰值電壓表,沖擊峰值電流表,沖擊峰值表,脈沖峰值電壓表,脈沖峰值電流表
設(shè)備簡介
NA23型沖擊峰值電壓表,用于測量和儲(chǔ)存沖擊電壓峰值、沖擊電流峰值以及其它單次脈沖峰值。內(nèi)部的一個(gè)計(jì)算單元可使分壓器高壓側(cè)的峰值以kV為單位直接顯示出來。顯示范圍從 0.1kV至 16000kV。
該裝置是為了高壓試驗(yàn)室的使用而專門設(shè)計(jì)的。采用特殊的布線、屏蔽、濾波以及過電壓保護(hù)線路使電磁干擾高度衰減。
測量范圍
通過測量電纜,該設(shè)備連接至沖擊電壓分壓器(電阻型或電容型設(shè)備)的低壓臂或大電流分流器上。對(duì)于測量分壓器(分流器), 沖擊峰值電壓表將引入一個(gè)2MΩ∥39pf的負(fù)載。如果需要,例如在電阻型沖擊分壓器的情況,可將75Ω(或50Ω)的匹配電阻加在該表的兩個(gè)輸入端之一。該裝置允許輸入的zui大脈沖電壓峰值為1600V。測量值以5位數(shù)字直接顯示,單位為kV。
該裝置不同于目前常用的沖擊峰值表的特點(diǎn)是:測量電壓量程無需手動(dòng)切換,而由裝置自動(dòng)切換。正負(fù)極性在顯示屏上自動(dòng)顯示。
主要參數(shù)
脈沖電壓顯示范圍:0.1kV~16000kV。
zui大脈沖電壓峰值為1600V。
電壓測量精度:誤差≤1%±0.1V。
分壓比設(shè)定: 0.1~9999.9。
采樣率選擇:40M/20M/10M/5M/2.5M/1.25M/0.6125M/0.30625M,對(duì)應(yīng)時(shí)間單位:1.5us/DIV,3us/DIV,6us/DIV,12us/DIV,24us/DIV,48us/DIV,96us/DIV,192us/DIV,384us/DIV,768us/DIV。建議4/10波形選擇3us/DIV,8/20波形選擇6us/DIV,2ms波形選擇384us/DIV。
或768us/DIV。
選擇400V量程時(shí),電壓刻度50V/DIV或25V/DIV或10V/DIV,儀表自動(dòng)設(shè)置;選擇20V量程時(shí),電壓刻度為2.5V/DIV或1.25V/DIV或0.5V/DIV。
外形尺寸: 490×130×300mm3。
重量: 4kg。
NA23 Type Impulse Peak-Voltmeter
Also name
NA23 impact peak voltage table, the impact pulse peak voltmeter, the impact of the peak ammeter, the impact of the peak table, the peak pulse voltage meter, pulse peak ammeter
Introduce of NA23
NA23 Impact Peak Voltmeter for measuring and storing impulse voltage peaks, peak impulse currents, and other single pulse peaks. An internal calculation unit allows the peak of the voltage divider to be displayed directly in units of kV. Display range from 0.1kV to 16000kV.
The tester designed for use with high pressure laboratories. Use of special wiring, shielding, filtering and over-voltage protection circuit to a high degree of attenuation of electromagnetic interference.
Measuring range
The measuring cable is connected to a low voltage arm or a high current shunt of an impulse voltage divider (resistive or capacitive type device). For measuring the voltage divider (shunt), the impulse peak voltmeter will introduce a load of 2MΩ // 39pf. If necessary, for example, in the case of a resistive shock divider, a 75Ω (or 50Ω) matching resistor can be added to one of the two inputs of the table. The maximum permissible pulse voltage of the device is 1600V. Measured values ??are displayed directly in 5 digits, in kV.
The device is different from the current commonly used impact peak table is characterized by: measuring voltage range without manual switching, and by the device automatically switch. Positive and negative polarity is automatically displayed on the display.
Main data of NA23
Pulse voltage display range: 0.1kV ~ 16000kV.
The maximum pulse voltage peak is 1600V.
Voltage measurement accuracy: error ≤ 1% ± 0.1V.
Partition ratio setting: 0.1 ~ 9999.9.
Sampling rate selection: 40M / 20M / 10M / 5M / 2.5M / 1.25M / 0.6125M / 0.30625M, corresponding time unit: 1.5us / DIV, 3us / DIV, 6us / DIV, 12us / DIV, 24us / DIV, 48us / DIV, 96us / DIV, 192us / DIV, 384us / DIV, 768us / DIV. Recommended 4/10 Waveform Select 3us / DIV, 8/20 Waveform Select 6us / DIV, 2ms Waveform Select 384us / DIV.
Or 768us / DIV.
When the 400V range is selected, the voltage scale is 50V / DIV or 25V / DIV or 10V / DIV. The instrument is set automatically. When the 20V range is selected, the voltage scale is 2.5V / DIV or 1.25V / DIV or 0.5V / DIV.
Dimensions: 490 × 130 × 300mm.
Weight: 4kg.